Philips Patient Monitoring Devices (Update C)

CVSS 6.8 MEDIUM

Risk Summary

Successful exploitation of these vulnerabilities could result in unauthorized access, interrupted monitoring, and collection of access information and/or patient data. To successfully exploit these vulnerabilities, an attacker would need to gain either physical access to surveillance stations and patient monitors or access to the medical device network.

Remediations

  • Philips released the following versions to remediate reported vulnerabilities:
  • Patient Information Center iX (PICiX) Version C.03
  • PerformanceBridge Focal Point
  • IntelliVue Patient Monitors Versions N.00 and N.01
  • IntelliVue Patient Monitors Version M.04: Contact a Philips service support team for an upgrade path
  • Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.
  • As a mitigation to these vulnerabilities, Philips recommends the following:
  • The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter.
  • By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices.
  • When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.
  • Implement physical security controls to prevent unauthorized login attempts on the PIC iX application. Servers should be kept in controlled locked data centers. Access to equipment at nurses' stations should be controlled and monitored.
  • Only grant remote access to PIC iX servers on a must-have basis.
  • Grant login privileges to the bedside monitor and PIC iX application on a role-based, least-privilege basis, and only to trusted users.
  • Users with questions regarding their specific Philips Patient Information Center (PIC iX) and/or IntelliVue patient monitor installations and new release eligibility should contact their local Philips service support team, or regional service support, or call 1-800-722-9377.
  • Please see the Philips product security website for the Philips advisory and the latest security information for Philips products.

Affected Vendors

Philips

Affected Products (9)

Philips · Patient Information Center iX (PICiX) B.02
Philips · Patient Information Center iX (PICiX) C.02
Philips · Patient Information Center iX (PICiX) C.03
Philips · PerformanceBridge Focal Point A.01
Philips · IntelliVue patient monitor MX100 <=N
Philips · IntelliVue patient monitor MX400-MX850 <=N
Philips · IntelliVue patient monitor MP2-MP90 <=N
Philips · IntelliVue X2 <=N
Philips · IntelliVue X3 <=N

Affected Sectors

Healthcare and Public Health